Description
System Function
- Throughput: 120 tests/hour
- Contents of testing: Clinical chemistry, Immunoturbidimetric
- Analytical modes: End-point, Two-point, Kinetic
- Programming: Open system with user defined programs
- Automatic, Random Access
- STAT sample priority
Sample Tray
- Sample position: 9 positions for samples
- Sample volume: 3 ~ 45 uL. 0.5 uL adjustable
- Any position for standard and quality control setting Reagent Tray
Reagent Tray
- Reagent position: 26 positions for reagents
- Refrigerated temperature: 4 – 15 C
- Reagent volume
R1: 180 ~ 450 uL, 1 pL adjustable
R2: 30 ~ 250 uL. 1 uL adjustable
Reaction Tray
- Reaction positions: 90 cuvettes, 10 cuvettes/strip
- Cuvette: Optical length 5mm
- Reaction volume: 180 – 500ul
- Reaction temperature: 37 C. fluctuation #0,4 C
Probe
- Pre-heating for reagent
- Collision protection, Liquid level detection and Inventory checking
- Automatic inner and outer probe washing after every cycle, carry-over<0.1%
Print Out
- Built-in thermal printer, external printer optional
Measuring and Optica System
- Lamp: Halogen lamp
- Absorbance range: 0 – 3.500Abs‡1%
- Wavelengths: 8 wavelengths from 340-700nm
- Resolution: 0.0001ABS